Is a Smart Nursing Bed Safe? What the Test Reports Actually Say
A smart nursing bed is electrical equipment used by a person who cannot get up on their own, so safety is not a marketing claim — it is a set of test reports with numbers. For the CH-ENB-01, the reports on file cover electrical safety, electromagnetic compatibility, the materials that touch the body, and the bed as a piece of furniture. This article lists each report, what it actually tested, and what it did not test. Read it before you believe any safety claim about any bed, including this one.
Start with the reports, not the adjectives
“Safe”, “certified” and “tested” are words any seller can print. Report numbers are different: they can be checked. A report has a number, a date, a standard it was tested against, and a result. If a seller cannot show you those four things, the claim is just an adjective.
The CH-ENB-01 has seven test reports on file, plus two attestations. Here they are, one by one, with what each one does and does not prove.
Electrical safety: the bed passed, and it must stay dry

Electrical safety was tested under report STE24011004S, against EN IEC 60335-1:2023+A11:2023 and EN 62233:2008. The result was: complied.
That matters because the bed is a 220–240V~, 50Hz, 950W appliance used inches from a person’s body, for indoor use only. The electrical safety test is the one that checks the bed will not give the user an electric shock in normal use.
But there is a boundary that belongs in the same sentence: the bed’s protection rating is IPX0. IPX0 means no protection against water ingress. The bed is not waterproof, and the whole bed must never be described as waterproof. Keep it dry, keep it indoors, and never hose it down or splash water over the electrical parts when cleaning. The automatic cleaning happens in a controlled part of the bed, with its own drainage — that is not the same as the bed being safe to get wet.
Electromagnetic compatibility: it will not disturb, or be disturbed
Electromagnetic compatibility was tested under report STE24011003E, against EN IEC 55014-1:2021, EN IEC 55014-2:2021, EN IEC 61000-3-2 and EN IEC 61000-3-3. The result was: emission Class B, overall pass.
In plain language: the bed’s electrical emissions are within the Class B limits for a home environment, and it passed the immunity tests. A home is full of other electrical equipment — phones, televisions, medical devices — and this test is the one that checks the bed plays quietly with them.
Materials: what the wood, fabric and sponge were tested for
Three reports from Chuangke cover the materials:
— Wood slats: formaldehyde release tested under GB 18584-2001, result 0.2 mg/L, pass — report CKJC2401071702-1.
— Fabric: formaldehyde content and 24 decomposable carcinogenic aromatic amine dyes tested under GB 18401-2010 Class B and GB/T 17592-2011, formaldehyde not detected, all 24 dyes not detected, pass — report CKJC2401071702-2.
— Sponge: formaldehyde content tested under GB 18401-2010 Class C, not detected, pass — report CKJC2401071702-3. Note the class: the sponge was tested to Class C, not Class B. Writing Class B for the sponge would be wrong.
These reports test the materials, not the finished bed as a whole. They are worth having, and they are worth describing exactly.
The whole bed as furniture: 16 items

The whole bed was tested as a piece of furniture under report SQ2306722, against QB/T 4190-2021. The verbatim conclusion is that 16 tested items met the testing reference. Formaldehyde release was 0.02 mg/m³ and TVOC was 0.047 mg/m³.
Write the number carefully: 16 items. Not 25. The English report SQ2306722 is the English version of the Chinese report SQ2305655, and the Chinese version prevails if there is any difference. This report is the one that looks at the bed as a finished product rather than as a set of materials — which is why its conclusion is worded the way it is: the tested items met the reference, item by item.
Biocompatibility: only two parts were tested
Biocompatibility was tested under report WT23080949, issued 2024-02-02, for batch 202309001 only. Only two skin-contact parts were tested: the urine baffle (silicone) and the toilet seat cushion (PU). The results were: in-vitro cytotoxicity grade 1 for the urine baffle and grade 2 for the toilet seat cushion, guinea-pig skin sensitization grade 0, and rabbit skin irritation PII 0.
Three things must be said plainly about this report. First, only those two parts were tested — the report does not cover the whole bed, and it must never be described as whole-bed biocompatibility testing. Second, the grades are 1 and 2 for cytotoxicity, not 0; writing grade 0 would be false. Third, the report gives no overall pass verdict. It reports the results of the tests it ran, and nothing more. Do not write “non-toxic” about the bed on the basis of this report.
The UDEM attestations: what they are, and what they are not
Two attestations from UDEM are on file, valid to 2029-01-14:
— EMC + LVD attestation M.2024.206.C95586, which cites EN 60335-1:2012+A15:2021 — a different version of the standard from the one cited in report STE24011004S, so cite each one separately.
— GPSD attestation M.2024.206.C95585.
An attestation is a statement of conformity, not an EU notified-body CE certificate. Calling either of these a notified-body certificate would be wrong. They support the product’s conformity documentation; they are not the same document as a certificate issued by a notified body.
What is not claimed
A careful safety page is also a list of things not claimed:
— The whole bed is not waterproof (IPX0).
— The whole bed was not biocompatibility tested — two skin-contact parts were, in one batch.
— The sponge was tested to Class C, not Class B.
— The furniture test covered 16 items, not 25.
— The UDEM documents are attestations, not notified-body certificates.
— No overall pass verdict exists for the biocompatibility report.
If a safety claim about any nursing bed cannot be traced to a report number, a standard and a result, treat it as a claim, not a fact. Ask the supplier for the report. Then read what the report actually says — including the parts that limit the claim.
How to read a report number
When you ask a supplier for evidence, ask for four things: the report number, the standard it was tested against, the date, and the result in the report’s own words. Then check that the number on the document matches the number in the answer, that the standard is the current one for the claim being made, and that the result is quoted exactly — “16 tested items met the testing reference” is a different claim from “passed 25 tests”, and only one of them is true.
For the CH-ENB-01, the numbers are: CKJC2401071702-1, CKJC2401071702-2, CKJC2401071702-3, SQ2306722, STE24011004S, STE24011003E and WT23080949, with UDEM attestations M.2024.206.C95586 and M.2024.206.C95585. Everything on this page can be checked against those documents.
Two more checks cost nothing. First, check the model number on the report matches the bed you are buying — a report for a different model proves nothing about this one, however impressive it looks. Second, ask for the full document, not a photograph of a wall of certificates or a thumbnail. A certificate wall shows that something was once certified; a report shows what was tested, against which standard, and with what result.
FAQ
Is the whole bed waterproof?
No. The protection rating is IPX0, which means no protection against water ingress. The bed is for indoor use only and must be kept dry.
Was the whole bed tested for biocompatibility?
No. Only two skin-contact parts were tested under report WT23080949 — the urine baffle (silicone, cytotoxicity grade 1) and the toilet seat cushion (PU, grade 2) — for batch 202309001 only, with no overall pass verdict.
What does the electrical safety report cover?
Report STE24011004S tested the bed against EN IEC 60335-1:2023+A11:2023 and EN 62233:2008, with the result: complied.
Are the UDEM documents CE certificates?
No. M.2024.206.C95586 and M.2024.206.C95585 are attestations of conformity, valid to 2029-01-14. They are not EU notified-body CE certificates.
How many items did the whole-bed furniture test cover?
16 tested items met the testing reference under report SQ2306722 (QB/T 4190-2021), with formaldehyde release 0.02 mg/m³ and TVOC 0.047 mg/m³.
See the full model, specifications and test reports on the product page. For a quote, please email admin@chinahospitalbed.com with your country, quantity and intended use.